Firmware Version 4.8.53 Release Notes 

Release Date: May 27, 2022 

Affected Models

TS029 - Compact RFID Refrigerator (Legacy)

TS031 - Midsize RFID Freezer (Legacy)

TS036 - Compact RFID Freezer (Legacy)

TS030 - Large RFID Cabinet

TS032 - Compact RFID Cabinet

TS035 - Compact RFID Cabinet (Europe)

TS047 - Compact RFID Freezer (Europe) (Legacy)

TS047e – Compact RFID Freezer (Europe)

TS056 - Large RFID Cabinet (Europe)

TS057 - Large RFID ULT Freezer (Legacy)

TS060 – Compact RFID Refrigerator (Europe)

TS079 - Compact RFID ULT Freezer

TS092 - Large Secure Access Cabinet

Affected Serial Numbers: All units with ZEUS SBCs that are currently running an earlier version of firmware v4.X

Firmware Version: 4.8.53 

Overview of changes: 

This is a bug release for field-installed units with ZEUS SBCs currently running an earlier version of firmware v4.X. Terso customer success team should deploy this firmware version to field units experiencing an I2C error.

Adds: 

  • Added retries into the I2C communication to make it more robust

Removes: 

None 

Updates: 

None 

Device ConfigParameters (Name, Description, Data Type, Min, Max, Default):

Adds:

None 

Removes:

None 

Changes:

None 

Device Definitions (List):

FW v4.08.052, Ambient Cabinet, Dup

FW v4.08.052, Standard Freezer, Dup

FW v4.08.052, Standard Refrigerator, Dup

FW v4.08.052, ULT, Dup