20221201 Enclosure FW v5.1.16 / OS 11.4.10 Release
Enclosure Firmware Version 5.1.16 Release Notes
Platforms: IRIS, IRIS Quad
Affected Models: This firmware (FW) update affects the following enclosure models with an IRIS or IRIS Quad SBCs that are currently running FW v5.0.XX.
- TS030 - Large RFID Cabinet
- TS032 - Compact RFID Cabinet
- TS035 - Compact International RFID Cabinet
- TS074e - Large RFID Refrigerator
- TS088 - Ultra Low Temperature RFID Freezer
- TS092 - Large Secure Access Cabinet
- TS093 - Midsize RFID Freezer
- TS096 - Undercounter RFID Freezer
- TS097 - Undercounter RFID Refrigerator
- TS098 - Countertop RFID Refrigerator
- TS099 - Midsize RFID Refrigerator
Firmware Version: FW v5.1.16 / OS 11.4.10
Application:
Production: FW v5.1.16 should be installed on all new production enclosure models with IRIS or IRIS Quad SBCs.
Field: FW v5.1.16 should be installed on all field-installed enclosure models with IRIS or IRIS Quad SBCs that are currently running v5.0.XX.
Overview of changes: This release adds new functionality and includes bug fixes.
Adds:
- Support for use of a barcode reader instead of the RFID pass reader for user access credentials (
- Support for Japanese language fonts for the graphics-capable LCD
- Added the ability to set the timeout on the DHCP client through the local interface and/or policy
- Additional debug messages during the AutoUpdate sequence
- New door lock functionality for the TS088. See Device Configuration Parameters for details.
Removes:
- N/A
Changes:
- A user of the local interface can now easily view the Interfaces and resolv.conf files to simplify the installation of devices in the field
- The timing of ObjectEvents is now a random second within the selected ObjectEventHour. Previously this was a random minute. The random second is automatically selected during the device boot sequence
- Improved memory management
- SBC Linux OS: Updated libraries
Known issues:
- NA
Device Configuration Parameters:
Adds:
- DoorLockTriggered – For TS088 only. When set to True, will enable DoorLockTriggeredDelay (see below) and the enclosure door lock is de-energized with a door breach.
- Type: Boolean
- Default
- DoorLockTriggeredDelay – For TS088 only. Time in seconds from when the door sensor picks up door closed to when door lock is energized. Only valid when DoorLockTriggered is set to true.
- Type: Integer
- Min: 1
- Max: 60
- Default: 3
Removes:
- N/A
Changes:
- N/A
Device Definitions:
Current device definitions for the previous version are still valid.
It is recommended that customers use the most current device definitions available unless instructed otherwise by the Terso Customer Success team.
- FW v5.1.xxx, Ambient Cabinet
- FW v5.1.xxx, Standard Freezer
- FW v5.1.xxx, Standard Refrigerator
- FW v5.1.xxx, ULT