Device Firmware Release Notes
...
Enclosure Firmware Releases
Firmware v4.x

20230104 Enclosure FW v4.8.55

5min

Platforms: Zeus, IRIS (pre-Firmware v5 only)

Affected Models: This firmware (FW) update affects the following enclosure models being built in production and field units currently running FW v4.X.

  • TS030 - Large RFID Cabinet
  • TS032 - Compact RFID Cabinet
  • TS035 - Compact International RFID Cabinet
  • TS047e - Compact International RFID Freezer
  • TS060 - Compact International RFID Refrigerator
  • TS074e - Large RFID Refrigerator
  • TS079 - Undercounter RFID Ultra Low Temperature Freezer
  • TS088 - Ultra Low Temperature RFID Freezer
  • TS092 - Large Secure Access Cabinet
  • TS093 - Midsize RFID Freezer
  • TS096 - Undercounter RFID Freezer
  • TS097 - Undercounter RFID Refrigerator
  • TS098 - Countertop RFID Refrigerator
  • TS099 - Midsize RFID Refrigerator
  • Any legacy models currently running FW v4.X

Firmware Version: FW v4.8.55

Application:

Production: FW v4.8.55 should be installed on all new production enclosures shipping with a Zeus SBC.

Field: FW v4.8.55 should be installed on all field-installed enclosure models with a Zeus or IRIS SBC that are currently running FW v4.8.53 or earlier.

Overview of changes: This release adds new functionality and changes to existing functionality.

Adds:

  • New door lock functionality for the TS088. See Device Configuration Parameters for details.

Removes:

  • N/A

Changes:

  • This version now uses the same language files as FW v5.X.
  • The timing of the start of ObjectEvents is now a random second within the selected ObjectEventHour. Previously ObjectEvents would start at the beginning of the selected hour. The random second is automatically selected during the device boot sequence.

Known issues:

  • NA

Device Configuration Parameters:

Adds:

  • DoorLockTriggered – For TS088 only. When set to True, will enable DoorLockTriggeredDelay (see below) and the enclosure door lock is de-energized with a door breach.
    • Type
      : Boolean
    • Default

  • DoorLockTriggeredDelay – For TS088 only. Time in seconds from when the door sensor state is "closed" to when door lock is energized. Only valid when DoorLockTriggered is set to true.
    • Type
      : Integer
    • Min
      : 1
    • Max
      : 60
    • Default
      : 3

Removes:

  • N/A

Changes:

  • N/A

Device Definitions:

It is recommended that customers use the most current device definitions available unless instructed otherwise by the Terso Customer Success team.

  • FW v4.8.55, Ambient Cabinet
  • FW v4.8.55, Standard Freezer
  • FW v4.8.55, Standard Refrigerator
  • FW v4.8.55, ULT