20230104 Enclosure FW v4.8.55
Platforms: Zeus, IRIS (pre-Firmware v5 only)
Affected Models: This firmware (FW) update affects the following enclosure models being built in production and field units currently running FW v4.X.
- TS030 - Large RFID Cabinet
- TS032 - Compact RFID Cabinet
- TS035 - Compact International RFID Cabinet
- TS047e - Compact International RFID Freezer
- TS060 - Compact International RFID Refrigerator
- TS074e - Large RFID Refrigerator
- TS079 - Undercounter RFID Ultra Low Temperature Freezer
- TS088 - Ultra Low Temperature RFID Freezer
- TS092 - Large Secure Access Cabinet
- TS093 - Midsize RFID Freezer
- TS096 - Undercounter RFID Freezer
- TS097 - Undercounter RFID Refrigerator
- TS098 - Countertop RFID Refrigerator
- TS099 - Midsize RFID Refrigerator
- Any legacy models currently running FW v4.X
Firmware Version: FW v4.8.55
Application:
Production: FW v4.8.55 should be installed on all new production enclosures shipping with a Zeus SBC.
Field: FW v4.8.55 should be installed on all field-installed enclosure models with a Zeus or IRIS SBC that are currently running FW v4.8.53 or earlier.
Overview of changes: This release adds new functionality and changes to existing functionality.
Adds:
- New door lock functionality for the TS088. See Device Configuration Parameters for details.
Removes:
- N/A
Changes:
- This version now uses the same language files as FW v5.X.
- The timing of the start of ObjectEvents is now a random second within the selected ObjectEventHour. Previously ObjectEvents would start at the beginning of the selected hour. The random second is automatically selected during the device boot sequence.
Known issues:
- NA
Device Configuration Parameters:
Adds:
- DoorLockTriggered – For TS088 only. When set to True, will enable DoorLockTriggeredDelay (see below) and the enclosure door lock is de-energized with a door breach.
- Type: Boolean
- Default
- DoorLockTriggeredDelay – For TS088 only. Time in seconds from when the door sensor state is "closed" to when door lock is energized. Only valid when DoorLockTriggered is set to true.
- Type: Integer
- Min: 1
- Max: 60
- Default: 3
Removes:
- N/A
Changes:
- N/A
Device Definitions:
It is recommended that customers use the most current device definitions available unless instructed otherwise by the Terso Customer Success team.
- FW v4.8.55, Ambient Cabinet
- FW v4.8.55, Standard Freezer
- FW v4.8.55, Standard Refrigerator
- FW v4.8.55, ULT